GE Healthcare – Lunar DPX IQFull Body Bone Densitometer, The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better. service manual for lunar dpx. Sat, 20 Oct GMT service manual for lunar dpx pdf – DPX Series. Service repair lunar dpx iq service manual – free. Frank’s Hospital Workshop. – service manual lunar dpx nt LUNAR DPX Operators . Manual, lunar dpx iq service manual pdf Gluer.

Author: Shaktijinn Kirn
Country: France
Language: English (Spanish)
Genre: History
Published (Last): 18 July 2004
Pages: 420
PDF File Size: 8.37 Mb
ePub File Size: 10.48 Mb
ISBN: 559-9-62706-948-1
Downloads: 44447
Price: Free* [*Free Regsitration Required]
Uploader: Mikagrel

You can adjust your Community Subscriptions in Settings. Contact Us Now – There’s lunqr thing i can do, and can someone send me the service manual, also?

Forums Documents Parts Videos News. Tissue value 1, Waiting for a prompt answer. This is a very common failure on any of the DPX models.

I have a problem with this team is that everything works, from the source to the Oink card, but does not open the solenoid of the collimator. The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better image quality. Tengo un problema con este equipo, resulta que todo funciona, desde la fuente, hasta la tarjeta Oink, pero no abre el solenoid del colimador.

  AGRIA 4800 PDF

service manual for lunar dpx

The language is no problem, actually is perfect. Hi Ben, I would like get you help. If active manually if the solenoid is activated, similarly, if I activate the switch, achievement cheat the system and I do not make the mistake, but if only the active from the system, it does nothing. Remove Subscription Confirm Remove.

Had the same problem was bad detector. If you would like a quote feel free to email me at ben dexasolutions. The DPX IQ provides the speed of entry level fan beam bone densitometer systems with better image quality.

Ad Statistics Times Displayed: Please review our Privacy Policy for more details. Just remember that the detector is not properly responsive during QC and exams as well.

Using a pencil beam to achieve the excellent precision and low radiation dose needed for the clinical environment. If active manually if the solenoid is activated, similarly, if I activate the switch, achievement cheat the system and I do not make the mistake, but if only the active from the system, it does nothing.

Remove Subscription Confirm Remove. You probably have a bad detector if the home positioning is good. My Bench Order History Sign out.

  GARABOMBO EL INVISIBLE MANUEL SCORZA PDF

service manual for lunar dpx

Greetings and thanks for your help Reply. If you like i find you a reloaded detector. On the momento the bmd is in 1, L2-L4,normal servive 1, to 1,the last reading before of the fail was 1, You may be getting improper results on the patient. Using a pencil beam to achieve the excellent precision and low radiation dose needed for the clinical environment.

Service manual for GE Lunar DPX-NT densitometer

We have it in stock if you need one. The detector is easy to replace but you will likely have to run the peak alignment. Greetings and thanks for your help Reply. Tengo un problema con este equipo, resulta que todo funciona, desde la fuente, hasta la tarjeta Oink, pero no abre el solenoid del colimador.

Dox too stock these detectors. You can sometimes get past this by using the x-ray test mode to saturate the detector.